RIE KLA SPTS Synapse
| RIE KLA SPTS Synapse | |
|---|---|
| Equipment Details | |
| Manufacturer | KLA |
| Model | SynapsEtch |
| Technology | RIE |
| Materials Restriction | Metals |
| Material Processed | GaN, Ga2O3, SiC, AlN |
| Mask Materials |
Photoresist SiO2 Si3N4 |
| Sample Size | 150 mm |
| Gases Used | Ar, O2, H2, Cl2, BCl3, SF6 |
| Equipment Manual | |
| Overview | System overview |
| Operating Procedure | SOP |
| Supported Processes | Supported Processes |
| User Processes | User Processes |
The SynapsEtch is an ICP RIE tool manufactured by KLA SPTS Technologies. The system is designed to etch strongly bonded materials such as GaN, AlN, SiC, sapphire, and Ga2O3 through use of a high density plasma source.
Contents
System overview
Hardware details
- Gases
- Ar
- O2
- H2
- Cl2
- BCl3
- SF6
- Pressure
- 2350 L/sec Pfeiffer ATH2804MT Turbo Pump
- 100 mTorr process gauge
- Fast acting VAT pendulum valve
- Chuck
- 150 mm Wafer
- 0-20 Torr Backside He Cooling
- -20°C to 55°C
- Chamber
- Aluminum Walls, Heated 55°C
- RF
- 3000 W, 13.56 Mhz Coil
- 1500 W, 13.56 Mhz Platen
Substrate Requirements
- 150 mm wafers
- 100 mm wafers and pieces can be mounted to a Si, SiO2 or Si3N4 carrier wafer with the Sample mounting
Material restrictions
The SynapsEtch is designated as a Metals class tool. A full list of approved materials is included at the end of this section. In addition to the restrictions in this list, materials can be classified into four categories, detailed below: materials that may be etched, materials that can be used as masks, etch stop materials, and buried materials. Use of any material outside of these conditions requires approval by the LNF staff via a helpdesk ticket.
Materials etched
This tool is intended for etching GaN, AlN, SiC, sapphire, and Ga2O3 and other III-V's.
Mask materials
This includes any material that will be exposed to the plasma for the majority of the process. The most common mask material is photoresist. SiO2 and Si3N4 can also be used. There are certain restrictions for where these materials may be deposited, as detailed in the approved materials list.
Etch stop materials
This includes any material that will be exposed briefly to the plasma. All materials listed in the approved materials list on the wiki are allowed, including approved mounting materials.
Buried materials
These materials may be present on the sample, but may not be exposed to the plasma. They may be covered by the mask or on the back of the sample, provided that the sample is mounted to a carrier wafer. Materials listed in the approved materials list on the wiki are allowed.
Approved materials
Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under the conditions described above. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.
Supported processes
Processes can be found here LNF_User:RIE_KLA_SPTS_Synapse_user_processes
For more information, please contact the tool engineers via the helpdesk ticket system.
Standard operating procedure
Checkout procedure
- Complete the sample mounting course. If you have already completed this for another tool, you do not need to complete it again.
- Read through this page and the User Manual above.
- Complete the process request form.
- Schedule a time for a training:
- Find an available time for checkout here.
- Verify that the tool is available in the LNF Online Services at the time you are requesting.
- After confirmation that the event is scheduled in Calendly, invite the staff member assigned to your event to a Staff Support reservation at that time.
- Create a helpdesk ticket for final confirmation of your checkout appointment.
- Schedule a time for a checkout using the above procedure.
- Authorization will be provided pending successful demonstration of proper tool use in the presence of a tool engineer.