|sop = [https://docs.google.com/a/lnf.umich.edu/document/d/1iRHIBHG0BvqpsX1quhb8YszCImOkyvzJx_SxqluPOt8/preview [{{PAGENAME}}#Standard_Operating_Procedure | SOP]|maintenance = [https://docs.google.com/a/lnf.umich.edu/document/d/1OvMzJ2QibI7rEU3ApebmoEPh_nvhBoVxd8NIxHiax0U/preview Maintenance Overview]
|processes = [[{{PAGENAME}}/Processes|Supported Processes]]
|userprocesses = [[LNF_User:Plasmatherm_790_User_Processes|User Processes]] }}
The Plasmatherm 790 is a dual chamber parallel plate tool. The left chamber is configured for [[Reactive ion etching|reactive ion etching (RIE)]] while the right chamber is configured for [[PECVD|plasma enhanced chemical vapor deposition (PECVD)]]. The RIE chamber is significantly slower than other etchers available in the lab, allowing greater depth control on thin films. The PECVD chamber is also designed to deposit at a lower rate than the [[GSI PECVD]] and [[P5000 PECVD]] - the smaller building blocks help improve film quality. This tool has almost no material restrictions, allowing a wide variety of processing.
==Announcements==
*[2014-09-25] - Added 500 sccm Helium MFC to PECVD chamber.
==Capabilities==
* RIE Gases
** [[Argon|Ar]] - 139 sccm
** [[Methane|CH<sub>4</sub>]] - 18 14 sccm
** [[Fluoroform|CHF<sub>3</sub>]] - 50 sccm
** [[Tetrafluoromethane|CF<sub>4</sub>]] - 42 sccm
* PECVD Gases
** [[Methane|CH<sub>4</sub>]] - 72 sccm
** [[Tetrafluoromethane|CF<sub>4</sub>- 4% O<sub>2</sub>]] - 420 sccm
** [[Helium|He]] - 500 sccm
** [[Nitrogen|N<sub>2</sub>]] - 2000 sccm
* Pressure
** RIE: 5 - 10 mTorr to 1 TorrGuage** PECVD: 50 - 500 mTorr to 2 TorrGauge
* Chuck
In addition to these, this tool has a number of user-created recipes for etching a wide variety of materials. Some of these recipes are documented on [[LNF User:{{BASEPAGENAME}} User Processes]]. If you are curious if your material can be processed in this tool, please contact the tool engineers via the helpdesk ticket system.
 
The latest PECVD qualification data is shown below.
 
{{#widget:Iframe
|url=https://docs.google.com/spreadsheets/d/1jDSV00jZ-1vppt2MgGRAJyXRvxEE3rJIlXtngODM9-0/pubhtml?gid=1394489689&amp;single=true&amp;widget=true&amp;headers=false
|width=650
|height=260
|border=1
}}
==Standard Operating Procedure==
* [https://drive.google.com/open?id=1ISeylrCbnE2g8E1acoAKEbNmKXi6CH2f System Overview]* [https://drive.google.com/file/d/0B3k-qgz56T_4RVJqUHp3eEsySUk/view Plasmatherm 790 ?usp=sharing&resourcekey=0-ClQBXI80fBwIL1FR48tt7w RIE SOP]* [https://drive.google.com/open?id=1iWQMtPGj8M9KQmTBJgTBk6jxuMiWCSk4 PECVD SOP]* [https://drive.google.com/open?id=1QiY9Xwt1ucw9axn4JTanLy-OA4QeXEIU Troubleshooting]
==Checkout Procedure==
<!-- Describe the checkout procedure for the tool. For example: -->
# Read through the [https://driveUser Manual above.google.com/file/d/0B3k-qgz56T_4RVJqUHp3eEsySUk/view Standard Operating Procedure]# Complete Accurately complete the [https://docs.google.com/a/lnf.umich.edu/forms/d/e/1FAIpQLSff-AcfmVpwlCTYKOQVe_sV0Rz_IE_B4vn0Pml7HX3eHcOC2w/viewform SOP quiz]. You may retake as necessary until all answers are correct.# Complete For RIE, complete the training request form [https://docs.google.com/a/lnf.umich.edu/forms/d/16kHlwOrAQk_nybzzXjYk78_aENmJiy33jJ7gj18lJKge/1FAIpQLSf4FcYP1XIoE-scXpXtmaegIzceSVe3uaR8e9P3K3LQAvn20w/viewform hereprocess request form].# Create a [http://ssel-sched.eecs.umich.edu/sselScheduler/ResourceContact.aspx?tabindex=3&path=0:0:0:{{#var:toolid}} helpdeskticket] ticket requesting traininga checkout session.# A tool engineer Authorization will schedule a time for initial training.# Practice with your mentor or another authorized user until you are comfortable with tool operation.# Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is be provided pending successful, completion of the engineer will authorize you on quiz and demonstration of proper tool use in the tool. ==Maintenance== Every other week presence of a light cleaning is performed on the toolengineer. A 5 min etch with [[Plasmatherm_790/Processes/m_si_o_1|m_si_o_1]] is performed on a 4" oxide wafer as a qualification run. The results can be seen below. {{#widget:Iframe|url=https://docs.google.com/spreadsheets/d/1jDSV00jZ-1vppt2MgGRAJyXRvxEE3rJIlXtngODM9-0/pubchart?oid=947675517&amp;format=interactive |width=700|height=350|border=0}}