The suggested starting parameters for bonding various materials are shown below. If you want to try bonding something else, we will have to characterize the bonding parameters.
Metal System | Wire Dia. | Work Holder temp. | Wedge Heater setting | 1st Bond | 2nd Bond | |||||
---|---|---|---|---|---|---|---|---|---|---|
Wire | Bond Pad | Force | Time | Power | Force | Time | Power | |||
Al | Al | .001" | Ambient | Off | 3 | 3 | 2.5 | 3 | 3 | 2.5 |
Al | Cu | .001" | Ambient | Off | 6 | 6 | 9 | 6 | 6 | 9 |
Al | Pt | .001" | Ambient | Off | 4 | 4 | 4 | 4 | 4 | 4 |
Au | Pt | .001" | 160°C | 36 | 4.5 | 4.5 | 4.5 | 4.5 | 4.5 | 4.5 |
Al | Au | .001" | Ambient | Off | 2.4 | 2.8 | 2.8 | 3.2 | 3.7 | 3 |
Au | Au | .0007" | 150°C | 36 | 3 | 3 | 0 | 3 | 3 | 0 |
Au | Au | .001" | Ambient | Off | 4 | 4 | 4 | 4 | 4 | 4 |
Au | Rough Au | .0007" | 150°C | 36 | 4.5 | 4.5 | 4.5 | 4.5 | 4.5 | 4.5 |
Al | Al | .0007" | Ambient | Off | 1.14 | 1.3 | 1.3 | 2.3 | 2.3 | 1.24 |
Au | Ag | .001" | Ambient | Off | 5 | 5 | 4 | 5 | 5 | 4 |