Difference between revisions of "AZ 12XT"

From LNF Wiki
Jump to navigation Jump to search
Line 21: Line 21:
 
* HMDS in YES Image Reversal Oven
 
* HMDS in YES Image Reversal Oven
 
**For Plating or glass RIE etching set the oven temp to 140°C. You need to allow 40 minutes each for the oven to heat and cool.
 
**For Plating or glass RIE etching set the oven temp to 140°C. You need to allow 40 minutes each for the oven to heat and cool.
**RIE etching can us HMDS applied at 90°C
+
**RIE and wet etching can us HMDS applied at 90°C
 
*Spin AZ 12xt
 
*Spin AZ 12xt
 
*Bake for 3 minutes on a 110°C hotplate with the lid lowered
 
*Bake for 3 minutes on a 110°C hotplate with the lid lowered

Revision as of 14:56, 1 October 2021

Warning Warning: This page has not been released yet.
About this Process
Process Details
Equipment AZ 12XT
Technology Lithography
Chemicals Used AZ 12XT, AZ 726



Az 12XT is a chemical enhanced positive tone photo resist (PR). In a chemical enhanced PR the exposure step creates a latent image and a week acid in the PR. The chemical reaction that makes the PR soluble in developer happens during the post exposure bake (PEB). This makes the length and temperature of the PEB extremely important. One of the advantages of the chemical enhance resits for thick resist is that it eliminate the waiting steps.

Procedure

for 10-17um layers

  • HMDS in YES Image Reversal Oven
    • For Plating or glass RIE etching set the oven temp to 140°C. You need to allow 40 minutes each for the oven to heat and cool.
    • RIE and wet etching can us HMDS applied at 90°C
  • Spin AZ 12xt
  • Bake for 3 minutes on a 110°C hotplate with the lid lowered
  • Exposure
    • 10-17um layers
      • Contact Alingers 12-16 sec
      • Stepper 0.27-0.3 sec, Focus -30-0.
    • >17um
      • Contact Alingers 20-22sec
  • Bake
    • For 3 and 4k spins - 1 minute on a 90° hotplate with the lid lowered
    • For a 2K spin - 1.5 minutes on a 90° hotplate with the lid lowered
    • For a 1K spin - 4 minutes on a 90° hotplate with the lid lowered
  • Develop on the CEE Develop stations using 726 DP 60-60
  • Hard Bake
    • For wet etching or platting hard bake at 110°C for 1 minute.
  • Removal
    • For thickness <17um spins removal as normal. (ash, PRS-2000, nanostrip)
    • For thickness >17um
      • PRS-2000 at 60°C for 10-15 minutes, or
      • Ash at temp lower than 100°C to break through RIE hardened layer
      • PRS-2000 after ash to remove most of PR

Characterization

Spin Curve

AZ 12xt spin curve on the CEE 100

Characterization Matrices

  • Exposure time vs Spin Speed on Contact Aligners

Az12xt 2-4K crosssections.jpg

  • Exposure time on Contact Aligners vs Development time

1k spin cross sections.jpg

  • Exposure time vs Focus on Stepper

EXPO cross section AZ12xt.jpeg

Data Sheets