Angstrom Engineering Evovac Evaporator

From LNF Wiki
Revision as of 13:07, 7 April 2017 by Davidsbn (talk | contribs)
Jump to navigation Jump to search

http://lnf-wiki.eecs.umich.edu/wiki/Evaporation

Warning Warning: This page is still under development Information is accurate but films will be added to charts as they are qualified.


Angstrom Engineering Evovac Evaporator
62061.jpg
Equipment Details
Technology PVD
Materials Restriction Metals
Material Processed Ti, Cr, Au, Al, Ni, Pt, Ag,Cu, Pd, Ge, Al2O3, SiO2, TiO2
Sample Size 6", 4", 3" and 2" wafers, pieces
Equipment Manual
Overview System Overview
Operating Procedure SOP
Supported Processes Supported Processes
Maintenance Maintenance



The Angstrom Engineering Evovac is a combination electron beam evaporator and thermal evaporator used to deposit Ti, Cr, Au, Al, Ni, Pt, Ag, Cu, Pd, Ge, Al2O3, SiO2 and TiO2 with the ability to heat the substrates and do in-situ substrate cleaning using an ion mill. It is an open loop box coater with a dome that can run multiple wafers and substrates in each run.

All the e-beam evaporators in the LNF are setup using small point sources and long throw distances that lead to very directional deposition with minimal heat transfer. The “line of sight” nature means that evaporated films have poor gap fill but make the tool ideal to use for lift-off applications. The evaporators in the LNF use thin films controllers with quartz crystal monitor feedback. These controllers pre-heat the source materials with the shutter closed. Then, during deposition, the controller uses the feedback from the crystal monitor and varies the e-beam or thermal power to maintain a steady, preset deposition rate and to determine the total evaporation thickness.

Announcements

  • May-June 2016 Tool being installed
  • 6/29/2016 SOP added, Data added for films so far. Tool will be released with a limited film set (10 ebeam, 2 thermal films)
  • 7/5/2016: Tool released to users with limited filmset.

Capabilities

  • Multi-wafer lift-off evaporator
    • Medium deposition rate (1-5 Å/sec) metal and dielectric films on multiple wafers
    • Designed for low heat transfer and minimal sidewall coverage for liftoff applications
    • Stacks of multiple films can be run

System Overview

Hardware Details

  • Cryo-pumped box coater capable of low 10-7 Torr base pressure in 1 hr of pumping
  • Reactive gasses or heating up to 250C.
  • 8x15cc pocket E-beam gun and 2 thermal sources to allow 10 different materials to be deposited in one run
  • Large distance from source and curved substrate dome designed for lift-off applications
  • Dome rotation and curved uniformity plate to allow for better uniformity

Substrate Requirements

  • 2”, 3”, 4” and 6” wafer holders available as well as clip fixtures for smaller pieces
  • Tool holds 4x4” or 4x6” wafers or 4x6" fixtures.

Material Restrictions

The Angstrom Engineering Evovac Evaporator is designated as a Metals class tool. Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.


Supported Processes

  • Angstrom Engineering Films:
Material E-beam Pocket E-beam Max thickness E-beam Max Deposition Rate Thermal Max thickness Thermal Deposition Rate
Titanium (Ti) 1 1 µm 5 Å/sec TBD TBD
Chromium (Cr) 2 5000 Å 5 Å/sec N/A N/A
Nickel Ni 3 5000 Å 5 Å/sec N/A N/A
Gold (Au) 4 1 µm 5 Å/sec TBD TBD
Aluminum (Al) 5 1 µm 5 Å/sec 2000 Å 2 Å/sec
Platinum (Pt) 6 2000 Å 5 Å/sec N/A N/A
Palladium Pd 6 5000 Å 5 Å/sec N/A N/A
Silver Ag 7 1 µm 5 Å/sec 2000 Å 2 Å/sec
Copper Cu 7 1 µm 5 Å/sec N/A N/A
Tantalum Ta 7 5000 Å 4 Å/sec N/A N/A
GermaniumGe 8 5000 Å 5 Å/sec N/A N/A
Alumina Al2O3 8 2000 Å 5 Å/sec N/A N/A
Silicon Dioxide SiO2 8 5000 Å 5 Å/sec N/A N/A
Titanium Dioxide TiO2 8 TBD TBD N/A N/A


  • Any combination of these films can be run as long as they are setup for the right pocket combination. They are run 1 at a time using the Inficon software.

Each of the materials has a standard Inficon film program outlined on the AE Evap Process Parameters page.

Standard Operating Procedure

Widget text will go here.

Checkout Procedure

  • Read through this page and the Standard Operating Procedure above.
  • Create a Helpdesk Ticket requesting training.
  • A tool engineer will schedule a time for initial training.
  • Practice with your mentor or another authorized user until you are comfortable with tool operation.
  • Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool.

Maintenance

The Angstrom Engineering Evaporator is opened for scheduled maintenance:

  • Once a week and the gun area is cleaned and shutter is replaced.
  • The chamber dome, sidewalls, shutters and covers are cleaned with foil replacement as needed (approximately once every 2-4 months.)
  • Tool leak checks, pumping speeds and film characterization are done periodically or when major tool changes are done.
  • Cryo absorbers are replaced once every 3 years.

Tool Pumping Performance

Pumping speeds and Leak-up rates vary based on chamber sidewall deposition and outside conditions such as humidity and time tool was open.

Angstrom Engineering pumping speed - tool crosses over to hi-vac in 6-7 minutes at 90mT

Process Characterization

Power Bump Recovery

The only maintenance that users are allowed to perform is power interruption recovery.