EVG 510
| EVG 510 | |
|---|---|
| Equipment Details | |
| Technology | Wafer bonding |
| Materials Restriction | Metals |
| Material Processed | Si, Glass |
| Sample Size | pieces, 100 mm, 150 mm |
| Equipment Manual | |
| Overview | System Overview |
| Operating Procedure | SOP |
| Supported Processes | Processes |
| Maintenance | Maintenance |
The EVG 510 is a wafer bonder capable of bonding pieces up to 150 mm wafers. It is primarily used for polymer bonding and wafer mounting with WaferBOND, but can also be used for other other types of bonding. Generally the EVG 520IS should be used for non-polymer bonding since the polymer bonding in EVG510 can leave contaminants that may interfere with non-polymer bonding. Also the CL200 Megasonic Cleaner can be used before bonding to reduce particulate.
Contents
Announcements
Please see the latest monthly update: Announcements
Starting 11/17/25 You will need check out the bond chuck. To do this, open a help desk ticket and request the required bond chuck. You will need to have a scheduled reservation on either the EVG 510 or EVG 520IS to check out a bond chuck. Your reservation parameters will have to match the allowable processes for the bond chuck and pressure plate combinations.
- Default: what is currently in the tool. If that is required in the EVG 520IS then the Old 4" Ti plate will be installed.
Use the drop down menus during reservation to indicate the tooling you will need. If the pressure plate it is different than what is on the calendar, open a help desk ticket to request a change. Include the type of bonding, sample size, pressures used, and when you will need this. Priority will be given to tickets with a corresponding tool reservation unless there are already reservations using the scheduled plate. You need to allow 2 business days before the next schedule change day. If there are existing reservations to use the scheduled plate, those will take precedence. Conflicting change request will be prioritized by when the request arrived and have corresponding tool reservations. If you don't have a tool reservation the request with a reservation will receive higher priority.
Capabilities
- Bonding
- Polymer
- Eutectic
- Wafer mounting via WaferBOND
System Overview
Hardware Details
- Force
- 0.1 to 10 kN
- Maximum pressure: 1.1 MPa (dependent on hardware and bonding method, see table below)
- Vacuum
- 1e-5 mBar
- ACP 28G Backing pump with Turbo
- Temperature
- 20°C to 550°C
- 45°C/min heating rate
- Passive cooling (1°C/min)
- Voltage
- ± 2000 V
- Up to 50 mA
Substrate Requirements
- Pieces, 100 mm, 150 mm
Material Restrictions
The EVG 510 is designated as a Metals class tool. Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.
Supported Processes
There are currently no characterized processes for the tool.
Maximum force
Bonding processes are characterized by a required pressure (force per area) in order to achieve bonding. The tool recipe specifies a force (in Newtons). This value should be scaled depending on the sample size to achieve the desired pressure. In addition, there is a maximum pressure that may be applied without damaging the equipment, which is dependent on the tooling material and whether the temperature is constant or varying while applying force.
The maximum pressure for the tooling available is shown in the following table. To calculate the force (in N), multiply the pressure (in MPa) by the sample area (in mm2).
| Pressure Disk Material | Pressure (MPa) | |
|---|---|---|
| Steady state | While heating | |
| Ti | 3.8 | 0.5 |
| Quartz | 0.6 | 0.2 |
Several common sample sizes are shown below. Note that small pieces cannot be used with the quartz plate, because they are below the minimum pressure that the tool can apply (100 N).
| Pressure Disk Material | Force (kN) | |||||||
|---|---|---|---|---|---|---|---|---|
| 150 mm wafer | 100 mm wafer | 50 mm wafer | 10x10 mm piece | |||||
| Steady state | While heating | Steady state | While heating | Steady state | While heating | Steady state | While heating | |
| Ti | 10 | 8 | 10 | 4 | 7.5 | 1 | 0.4 | N/A |
| Quartz | 10 | 5 | 5 | 2 | 1.25 | 0.5 | N/A | N/A |
Standard Operating Procedure
Widget text will go here.
Checkout Procedure
- Read through this page and the Standard Operating Procedure above.
- Create a Helpdesk Ticket requesting training.
- A tool engineer will schedule a time for initial training.
- Practice with your mentor or another authorized user until you are comfortable with tool operation.
- Complete the SOP quiz here.
- Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool.