EVG 520IS

From LNF Wiki
Jump to navigation Jump to search



EVG 520IS
140041.jpg
Equipment Details
Technology Wafer bonding
Materials Restriction Metals
Material Processed Si, Glass
Sample Size pieces, 100 mm, 150 mm
Equipment Manual
Overview System Overview
Operating Procedure SOP
Supported Processes Processes
Maintenance Maintenance


The EVG 520 is a wafer bonder capable of bonding pieces up to 150 mm wafers. It is primarily used for anodic, fusion and thermocompression bonding. Also the CL200 Megasonic Cleaner can be used before bonding to reduce particulate.

All polymer bonding should be performed in the EVG 510

Announcements

Pressure Plate Schedule

Unless otherwise specified, the default pressure plate will be installed in the tool. For all other pressure plate change requests, please submit a helpdesk ticket at least 2 business days in advance. If there are existing reservations to use the scheduled plate, those will take precedence.

  • Default: 6" Ti
Users are not authorized to change pressure plates.

Capabilities

  • Bonding
    • Anodic
    • Eutectic
    • Thermocompression
    • Fusion

System Overview

Hardware Details

  • Force
    • 0 to 60 kN
    • Maximum pressure: 3.5 MPa (dependent on hardware and bonding method)
  • Vacuum
    • 1e-5 mBar
    • ACP 28G Backing pump with Turbo
  • Temperature
    • 20°C to 550°C
    • 45°C/min heating rate
    • Active cooling (?°C/min)
  • Voltage
    • ± 2000 V
    • Up to 50 mA

Substrate Requirements

  • Pieces, 100 mm, 150 mm

Material Restrictions

The approved materials listed for the EVG 520IS are restricted to those that can come into surface with the tooling. There are several exceptions that are allowed within the bond interface, depending on the type of bond used. Some of these can be found in the #supported processes section. For any other materials, please create a helpdesk ticket.

The EVG 520IS is designated as a Metals class tool. Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.


Supported Processes

The supported processes for this tool are described in detail in the supported processes document. For a specific process the maximum chuck force should not exceed the guidelines below.

Pressure Disk

Material

Force (kN)
150 mm 100 mm 75 mm 50 mm
Steady

State

While

Heating

Steady

State

While

Heating

Steady

State

While

Heating

Steady

State

While

Heating

Ti 60 8 30 4 1.5 1.5 6 1
Quartz 10 5 5 2 3 1 2 0.5

Standard Operating Procedure

Widget text will go here.

Checkout Procedure

  1. Read through this page and the Standard Operating Procedure above.
  2. Create a Helpdesk Ticket requesting training.
  3. A tool engineer will schedule a time for initial training.
  4. Practice with your mentor or another authorized user until you are comfortable with tool operation.
  5. Complete the SOP quiz here.
  6. Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool.

Maintenance

Once a week the chucks are cleaned and inspected for damage. A monitor run is performed every other month on a 150 mm wafer to ensure that the tool is in good operating condition.