Energy Dispersive X-ray Spectroscopy (EDS, EDX) or Energy Dispersive Analysis is an analytical technique that can give quantitave and qualitative information on the elements that are present in an specimen. EDS takes advantage of the x-ray generated on a sample by the high energy electron beam of the Scanning Electron Microscope (SEM). Each atom releases x-rays whose finger print is unique.
Contents
Equipment
Both technique use the same exciting source and are housed in the same equipment.
Method of operation
The electron beam (>15keV) mainly generates two types of excited electrons: secondary and primary backscattered. As secondary electrons leave the sample the specimen charges (ionizes) and, in order to relax (neutralize) the atoms, a higher level electron moves into a lower level. This decay results in energy being released as photons (x-rays). These x-rays are a result of energy level transition within the energy levels of the atom, therefore the energy liberated is characteristic of each atom.
E=hc/ʎ̩, E= Ehigh-Elow
It is this characteristic energy that can then be associated to the electromagnetic radiation emitted, the x-rays. The x-rays intensity are measured versus their energy (dispersive).
Parameters
Electron Beam Energy: this must be high enough to excite core electrons. It is recommended to use at least 2.3x the x-ray energy to get sufficient signal for the element you are trying to detect.
Atomic number of the elements: EDS works better for higher Z elements
Sample preparation: sample should be flat and clean
Similar elements, elements that are close in the periodic table, or belong to the same group, often have overlapping peaks which difficult their identification
Subtechnology
Both techniques take place in the same instrument, use same exciting source, but different detectors.
Applications
This technique is a very useful tool for the identification and quantification of the elements present in a sample. It relies on the SEM highly energetic beam to create x-ray emission in a sample, therefore it is not suitable for thin films <100nm analysis; it is also not recommended for the analysis of low z elements. This technique does not give oxidation state information i.e. elemental metal and metal oxide both contribute to the metal atom signal and are undistinguishable.
See also
Other related wiki pages
References
- "Scanning Electron Microscopy and X-ray Microanalysis", 3rd Ed.,Goldstein, Newbury, Joy, Lyman, Echlin, Lifshin and Sawyer, 2003, Plenum Publishing Corp., ISBN13: 978-0306472923
- “Transmission Electron Microscopy: A Textbook for Materials Science,” 2nd Ed., D. B. Williams and C. B. Carter, Springer, New York, 2009, pp. 1-757