Difference between revisions of "LAM 9400/Processes/LNF Poly"

From LNF Wiki
Jump to navigation Jump to search
 
(13 intermediate revisions by the same user not shown)
Line 10: Line 10:
 
[[Category:Processes]] [[Category:RIE]]
 
[[Category:Processes]] [[Category:RIE]]
  
LNF_Poly is designed for etching si and polysilicon in the LAM 9400.  This recipe is currently under development and will replace poly_etch_2 as it does not have the same stability problems.
+
LNF_Poly is designed for etching [[Silicon|Si]] and [[Polysilicon]] in the LAM 9400.  This recipe produces a very vertical sidewall angle and has a high selectively to oxide (~18:1).
  
 +
==Etch Rates==
 +
* Polysilicon - 2340 Å/min (blanket 150mm poly)
 +
*Si - 2800 Å/min (<10% open area)
 +
*SPR 220 (3.0) - 275 Å/min
 +
*SiO<sub>2</sub> - 150 Å/min
 +
 +
{{#widget:Iframe
 +
|url=https://docs.google.com/spreadsheets/d/1fMQ3FPo6sf2o-e0B2FYGRTN6iBklOu8UjS5TBbYDg70/pubchart?oid=1633184043&format=interactive
 +
|width=600
 +
|height=400
 +
|border=0
 +
}}
  
 
==Parameters==
 
==Parameters==
Line 20: Line 32:
 
! Main Etch
 
! Main Etch
 
|-
 
|-
| Pressure
+
| Time
 +
| 10 sec
 +
| variable
 +
|-
 +
|Pressure
 
| 5 mTorr
 
| 5 mTorr
 
| 30 mTorr
 
| 30 mTorr
Line 45: Line 61:
 
|}
 
|}
  
==Capabilities==
+
==Sidewall Profile==
===Etch Rate===
+
* Nearly a 90 degree sidewall angle
 
 
===Mask Selectivity===
 
  
===Sidewall Profile===
+
<gallery mode="packed-hover"  heights="300px">
[[File:LAM_9400_LNF_Poly_2um_grating.jpg|400px]]
+
File:LAM_9400_LNF_Poly_2um_grating.jpg|LNF_Poly for 600 sec, 2 µm grating
 +
File:LAM_9400_LNF_Poly_10um_grating.jpg|LNF_Poly for 600 sec, 10 µm grating
 +
</gallery>
  
 
==Limitations==
 
==Limitations==
 +
To remove the photoresist a ~20 sec dip in 100:1 HF is necessary before nanostrip.

Latest revision as of 11:41, 20 October 2021


About this Process
Process Details
Equipment LAM 9400
Technology RIE
Material Poly
Mask Materials PR
Gases Used HBr, He
Date Created 2016

LNF_Poly is designed for etching Si and Polysilicon in the LAM 9400. This recipe produces a very vertical sidewall angle and has a high selectively to oxide (~18:1).

Etch Rates

  • Polysilicon - 2340 Å/min (blanket 150mm poly)
  • Si - 2800 Å/min (<10% open area)
  • SPR 220 (3.0) - 275 Å/min
  • SiO2 - 150 Å/min

Parameters

Parameter Breakthrough Main Etch
Time 10 sec variable
Pressure 5 mTorr 30 mTorr
TCP Power 300 W 600 W
Bias Power 100 W 90 W
Cl2 Flow 60 sccm 0 sccm
HBr Flow 0 sccm 100 sccm
He Flow 0 sccm 100 sccm

Sidewall Profile

  • Nearly a 90 degree sidewall angle

Limitations

To remove the photoresist a ~20 sec dip in 100:1 HF is necessary before nanostrip.