LAM 9400/Processes/LNF TiN sel Al2O3

From LNF Wiki
< LAM 9400‎ | Processes
Revision as of 09:30, 2 July 2019 by Wrightsh (talk | contribs) (Created page with "<!-- Make sure to add any other relevant categories --> {{Infobox process |technology = RIE |material = TiN, Ti |mask = Photoresist, SiO<sub>2</sub>, Al<sub>2</sub>O<sub>3</su...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search


About this Process
Process Details
Equipment LAM 9400
Technology RIE
Material TiN, Ti
Mask Materials Photoresist, SiO2, Al2O3
Gases Used Ar, Cl2, SF6
Date Created June 2019
Authored By Shawn Wright

This recipe is designed to etch TiN while being very selective to Al2O3. We currently don't have a clear etch rate for the TiN, however it's >100nm/min.

Etch Profile

Etch Rates

These were measured using 1cm pieces on a 6" Si carrier.

Main Etch

  • TiN: >150 nm/min
  • Veeco Al2O3: 3.2 nm/min
  • SiO2: 46.9 nm/min


Parameters

Parameter Main Etch
Pressure 10 mTorr
TCP Power 400 W
Bias Power 20 W
Vpk-pk ~30 V
Ar Flow 84 sccm
Cl2 Flow 36 sccm
SF6 Flow 10 sccm
Time 10-60 sec (typ)