Difference between revisions of "Lab 18-1"

From LNF Wiki
Jump to navigation Jump to search
Line 107: Line 107:
 
==Standard Operating Procedure==
 
==Standard Operating Procedure==
 
<!-- To include a document from google docs, use the line below, replace "googledocid" with the ID for the document. Remember, to make this visible, you must set Sharing for the document to "Anyone with the link can view". -->
 
<!-- To include a document from google docs, use the line below, replace "googledocid" with the ID for the document. Remember, to make this visible, you must set Sharing for the document to "Anyone with the link can view". -->
{{#widget:GoogleDoc|key=1JZafXTZbvQkeqPWAHTm_LTwAPXk7_8Su9Q-swt_I1xQ}}
+
{{#widget:GoogleDoc|key=1R_uj2oR6XJHc4D-J8opSyOfX-Xjf9-9ZVN3Jsln4c_o}}
  
 
==Checkout Procedure==
 
==Checkout Procedure==

Revision as of 09:49, 12 November 2021

< Sputter deposition

Lab 18-1
61071.jpg
Equipment Details
Technology PVD
Materials Restriction Metals
Material Processed Al2O3, ITO, Mo, Si, SiO2, Si3N4, Ta, Ta2O5, Ti
Sample Size 150mm x 4mm and smaller samples
Gases Used Ar, N2, O2
Equipment Manual
Overview System Overview
Operating Procedure SOP
Supported Processes Supported Processes
User Processes User Processes


Lab 18-1 is a magnetron sputter deposition tool for depositing less common metals, insulators, and semiconductive films. Sputter deposition is achieved by bombarding a source material with energetic ions, typically Ar+. Atoms at the surface of the target are knocked loose, and transported to the surface of the substrate, where deposition occurs. Electrically conductive material such as Al, W, and Ti can use a DC power source, in which the target acts as the cathode in a diode system. Sputtering of dielectrics such as silicon dioxide, or aluminum oxide requires an RF power source to supply energy to the argon atoms. Sputter films are ideal for sidewall coverage in low thermal budget cases for contact and insulating layers over features.

Announcements

Capabilities

  • Computer controlled recipes
  • Load lock supported processing
  • Limited film stress versatility with in film development.
  • Reactive film capabilities (Ar, O2, and N2)

Supported Processes

Basic "User" access to the Lab 18's provide researchers with the ability to load, transfer samples, and deposit standard characterized materials. For more details, see Lab 18-1/Processes.

In addition to these, this tool has a number of user-supported recipes for depositing a wide variety of materials. Some of these recipes are documented on User Processes.

If you have a material deposition need that is not listed, check to see if this material is deposited on on another PVD tool on the LNF PVD Films page, or see more options for running your own private material on the Target configuration section of the Lab 18-1/Processes page. The PVD 75 Proline is the preferred sputtering tool for non-standard materials. See how to Request New PVD Films here. See more on how to request new sputter sources specifically on the LNF Sputter Adding New Films page.

Sputter Materials

Seven films are available. If two thicknesses are listed the first is for per run and the second thickness is for total material used per reservation.

Material Max thickness (Run/Reservation)
Alumina/Aluminum Oxide (Al2O3) 100Å / NA
Indium-Tin-Oxide (ITO) 500 Å / 3000 Å
Molybdenum (Mo) 3000 Å / 8000 Å
Silicon (p-doped) (Si) 3000 Å / 1 µm
Silicon dioxide (SiO2) 1000 Å / 3000 Å
Silicon Nitride (Si3N4) 500 Å / 2000 Å
Tantalum (Ta) 3000 Å / 1 µm
Titanium (Ti) 2000 Å / 6000 Å

System Overview

The Lab 18-1 has a 18" D-shaped chamber with a load lock chamber to quickly load and unload samples without breaking high vacuum in the process chamber. The tools five magnetron sputtering sources are set up to allow for RF, DC, and DC co-sputtering. There is an additional RF power supply on the substrate platen to allow for in situ sample cleans, and bias voltage to manipulate the film characteristics. Sample heating up to 500°C is achieved by using the tools quartz heater lamps.

Hardware Details

  • Turbo pumped chamber – upper 10-7 Torr base pressure
  • Chamber capacity: single wafer
  • Configured with load lock
  • Sample heating – operable up to 450°C
  • Power specs
    • Five 3” Sputtering Guns
    • 2 DC and 2 RF power supply (one for sample bias)

Substrate Requirements

  • Sample sizes: pieces, up to 6” wafers.
  • Wafer holders available for 2", 3", 4", 6"; as well as clip fixtures for smaller pieces
    • Diameter 150 mm maximum
    • Substrates up to 4 mm thick, with special fixture up to 8 mm.
    • Sample Mounting

Material Restrictions

The Lab 18-1 is designated as a Metals class tool. Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.


Standard Operating Procedure

Widget text will go here.

Checkout Procedure

For more details, see KJLC Sputter Tool Access and Training.

  1. Complete the online PVD Rise Training (this only needs to completed once so if you are authorized on another PVD tool or sputter tool, you do not need to do it again.)
  2. Read through this Wiki page and the Operating Procedure above.
  3. Create a Helpdesk Ticket requesting training
  4. A tool engineer or user services member will contact you for a training session.
  5. Practice with your mentor or another authorized user until you are comfortable with tool operation.
  6. Complete the Lab 18 User Training Quiz
  7. Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool at the "user" level.

Maintenance

The sputter shield cleaning procedure is provided below. The source shields are stored on the rack between the Lab 18s when they are not installed. Some films that build up on the shield risk flaking off and shorting out plasma (W-Ti and ITO are the most frequent offenders). This procedure mitigates that, so it is posted here to empower users.

Lab 18 Shield Cleaning Procedure

For more details, see Lab 18-1/Maintenance and qualification.