Difference between revisions of "MA6 Mask Aligner"
Line 17: | Line 17: | ||
|maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]] | |maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]] | ||
}} | }} | ||
− | + | The [[{{PAGENAME}}]] is primarily used to expose [[Photoresist|photo-definable]] materials. We allow 4” masks, 5” masks, 7” masks, transparencies, shadow masking, or just flood exposure. | |
− | |||
==Capabilities== | ==Capabilities== | ||
<!--A more general description of what the tool is capable of doing.--> | <!--A more general description of what the tool is capable of doing.--> |
Revision as of 15:03, 6 November 2015
MA6 Mask Aligner | |
---|---|
Equipment Details | |
Technology | Contact Lithography |
Materials Restriction | Semi-Clean |
Material Processed | Photoresist |
Mask Materials | 4", 5", 7" mask plates |
Sample Size | 4" and 6" wafers. Pieces with staff approval |
Chemicals Used | LNF approved photoresists |
Equipment Manual | |
Overview | System Overview |
Operating Procedure | SOP |
Supported Processes | Supported Processes |
Maintenance | Maintenance |
The MA6 Mask Aligner is primarily used to expose photo-definable materials. We allow 4” masks, 5” masks, 7” masks, transparencies, shadow masking, or just flood exposure.
Contents
Capabilities
- 1-2 µm min feature size resolution in thin resist
- Aspect Ratio 1:3 fairly standard
- Substrate thickness 200 µm to 4 mm possible
Announcements
- Update this with announcements as necessary
System overview
Hardware Details
- Broadband light source calibrated to 20 mJ/sec at the 405 nm peak. Total broadband dose about 30 mJ/sec
- 4" and 6" chucks
- 4", 5" and 7" mask holders
- Flood exposure, soft, hard, and vacuum contact capable (for details, see SOP)
- Capable of topside alignment only
- 10 mm stage travel, 150 mm microscope objective travel with split field view camera option
Substrate requirements
- Full 4" and 6" wafers standard
- Pieces possible with tool engineer approval
- Mounting may be necessary if wafer has thru holes, etc. Create helpdesk ticket for assistance
- Substrate thickness 200 µm to 4 mm possible
Material restrictions
The MA6 Mask Aligner is designated as a Semi-Clean class tool. Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.
Supported processes
This tool is primarily used to expose photodefinable materials. The lamp output is calibrated to the 405 nm peak at 20 mJ/sec with 3% uniformity across the 4” area. The 365 peak output is about half of this and tracks very well however is not routinely calibrated. We allow 4” masks, 5” masks, 7” masks, transparencies, shadow masking, or a manually timed flood exposure.
Exposure of all LNF approved photoresists are allowed on this tool in accordance with standard operating procedure. More data on exposure times can be found on datasheet for the specific photoresist.
In addition to these, this tool has a number of user-created recipes for etching a wide variety of materials. Some of these recipes are documented on MA6 Mask Aligner user processes. If you are curious if your material can be processed in this tool, please contact the tool engineers via the helpdesk ticket system.
Standard operating procedure
Widget text will go here.
Checkout procedure
- Read through this page and the Standard Operating Procedure above.
- Create a Helpdesk Ticket requesting training.
- A tool engineer will schedule a time for initial training.
- Practice with your mentor or another authorized user until you are comfortable with tool operation.
- Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool.
Maintenance
The UV energy at the chuck face is routinely calibrated, so that the energy delivered corresponds to the display on the lamp controller. The energy delivered is 20 mJ/sec at the 405 nm wavelength with an additional 9-10 mJ/sec at the 365 nm wavelength (the 365 nm peak is not calibrated, but is constant in relation to the 405 nm peak). Lamp uniformity is routinely calibrated to within 2-3% across the 150 mm range. The WEC (Wedge Error Compensation) head is completely rebuilt semi annually and checked monthly and adjusted as needed for level.