Open main menu

< Sputter deposition

PVD 75 Proline
87001.jpg
Equipment Details
Technology PVD
Materials Restriction Metals
Material Processed Al, Al2O3, Cr, Cu,Co,Fe, Ni,Mo, Si, SiO2, Si3N4, Ag,Ti,W
Sample Size 150mm x 4mm and smaller samples
Gases Used Ar, N2, O2
Equipment Manual


The PVD 75 Proline is a magnetron sputter deposition tool for depositing conductive metallic and non conductive dielectric films. The PVD 75 tool is designed specifically for point of use processing of materials not allowed or not currently in the Lab 18 tools. Sputter deposition is achieved by bombarding a source material with energetic ions, typically Ar+. Atoms at the surface of the target are knocked loose, and transported to the surface of the substrate, where deposition occurs. Electrically conductive material such as Al, W, and Ti can use a dc power source, in which the target acts as the cathode in a diode system. Sputtering of dielectrics such as silicon dioxide, or aluminum oxide requires an Rf power source to supply energy to the argon atoms. Sputter films are ideal for sidewall coverage in low thermal budget cases for contact and insulating layers over features.

Contents

Announcements

Capabilities

  • Materials for deposition can be changed each reservation
  • Samples loaded directly into the process chamber
  • Oxygen (O2) and Nitrogen (N2) reactive film capabilities

Supported Processes

Basic "User" access with the Lab 18's provide researchers the ability to load, transfer samples, and deposit standard characterized materials. For more details, see PVD 75 Proline/Processes.

If you have a material deposition need that is not listed, check to see if this material is deposited on another PVD tool or see more about options for running your own private material on the LNF PVD Films page. The PVD 75 Proline is the preferred sputtering tool for non-standard materials. See how to Request New PVD Films here. See more on how to request new sputter sources specifically on the LNF Sputter Adding New Films page.

Process Name

Multiple staff characterized films are available. If two thicknesses are listed the first is for per run and the second thickness is for total material used per reservation.

Material Max thickness (Run/Reservation)
Aluminum (Al) 3000Å / 1 µm
Alumina/Aluminum Oxide (Al2O3) 500Å / 3000Å
Chromium (Cr) 3000Å / 9000Å
Iron (Fe) 2000Å / 6000Å
Nickel (Ni) 2000 Å / 6000Å
Molybdenum (Mo) 3000 Å / 8000 Å
Silver (Ag) 3000 Å / 6000Å
Silicon (Si) 3000 Å / 6000Å
Silicon Dioxide (SiO2) 1000 Å / 3000Å
Silicon Nitride (Si3N4) 500 Å / 1000Å
Titanium (Ti) 3000 Å / 6000 Å
Tungsten (W) 3000 Å / 9000 Å

System Overview

The PVD 75 is an Argon sputter deposition tool that uses Torus Mag Keeper magnetrons powered by RF, DC, and Pulsed DC power supplies. Deposition rates are between 0.1 and 11 A/sec, with most being close to 2 A/sec. An additional Rf power supply on the substrate platen allows for in situ sample cleans, and bias voltage. Quartz heater lamps behind the sample holder can be use to heat samples up to 300°C. The average run time on this tool per wafer is about two hours.

Hardware Details

  • Cryo pumped chamber – lower 10-6 Torr base pressure
  • Chamber capacity: single wafer
  • Sample heating – up to 300°C
    • Four 3” Sputtering Guns
    • 2 DC (one pulse DC source) and 2 RF power supply (one for sample RF bias)

Substrate Requirements

  • Sample sizes: pieces, up to 6” wafers.
  • Wafer holders available for 2", 3", 4", 6"; as well as clip fixtures for smaller pieces
    • Diameter 150 mm maximum
    • Substrates up to 4 mm Thick. With special fixture up to 8 mm.
    • Sample Mounting

Material Restrictions

The PVD 75 Proline is designated as a Metals class tool. Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.


Standard Operating Procedure

Widget text will go here.

Checkout Procedure

For more details, see KJLC Sputter Tool Access and Training.

  1. Complete the online PVD Rise Training (this only needs to completed once so if you are authorized on another PVD tool or sputter tool, you do not need to do it again.)
  2. Read through this Wiki page and the Operating Procedure above.
  3. Create a Helpdesk Ticket requesting training
  4. A tool engineer or user services member will contact you for a training session.
  5. Practice with your mentor or another authorized user until you are comfortable with tool operation.
  6. Complete the PVD 75 User Training Quiz
  7. Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool at the "user" level.

Maintenance