Difference between revisions of "Plasma etching"

From LNF Wiki
Jump to navigation Jump to search
 
(19 intermediate revisions by 3 users not shown)
Line 6: Line 6:
 
}}
 
}}
  
Plasma etching is a form of [[plasma processing]] designed to [[etching|remove]] material from a sample using plasma discharges.  It is highly controllable and can be used to etch a wide variety of materials.  The most commonly used form of plasma etching is referred to in the microfabrication world as [[reactive ion etching]] (RIE).  However, there are other types of plasma etching, including [[plasma ashing]] and [[ion milling]].  For a detailed overview of plasma etching in the LNF, please review the [https://docs.google.com/a/lnf.umich.edu/file/d/0B76AgohVTgqdamFiaE1mRk9mVmM/preview technology workshop]
+
Plasma etching is a form of [[plasma processing]] designed to [[etching|remove]] material from a sample using plasma discharges.  It is highly controllable and can be used to etch a wide variety of materials.  The most commonly used form of plasma etching is referred to in the microfabrication world as [[reactive ion etching]] (RIE).  However, there are other types of plasma etching, including [[plasma ashing]] and [[ion milling]].  For a detailed overview of plasma etching in the LNF, please review the [https://drive.google.com/file/d/0B76AgohVTgqdamFiaE1mRk9mVmM/preview?usp=sharing&resourcekey=0-yGk7A8kAmb0kQArcKPNlTQ Tech Talk]
 +
 
 +
{{TOC|limit=2|clear=left}}
 +
 
 +
==Reactive ion etching==
 +
{{main|Reactive ion etching}}
 +
 
 +
Reactive ion etching (RIE) is one of the most common forms of [[plasma etching]].  It typically uses a combination of chemically reactive elements and energetic ions to etch the desired material.  One major advantage to RIE over other forms of etching is that the process can be designed to be highly [[Isotropy|anisotropic]], allowing for much finer resolution and higher aspect ratios.
 +
 
 +
{{:Reactive ion etching}}
 +
==Deep reactive ion etching==
 +
{{main|Deep reactive ion etching}}
 +
 
 +
Deep reactive ion etching (DRIE) is a term used to describe RIE processes that are designed to achieve high (>10:1) aspect ratios or etch depths greater than several microns. One common process is the [[Wikipedia:Bosch process|Bosch process]].
 +
 
 +
{{:Deep reactive ion etching}}
 +
==Plasma ashing==
 +
{{main|Plasma ashing}}
 +
 
 +
Plasma ashing typically refers to the removal of organics, particularly [[photoresist]] from a sample using a plasma discharge.  These processes typically use [[oxygen]] as the main etch gas and sometimes require a high temperature to enhance the reactivity.
 +
 
 +
===YES-CV200RFS(E)===
 +
The [[YES-CV200RFS(E)|YES-CV200RFS(E) (YES Plasma Stripper)]] is the primary plasma ashing equipment in the main cleanroom of the LNF.  Its primary uses are low-temperature [[plasma descum]] and high-temperature [[photoresist stripping|stripping]] of [[photoresist]] but can also be used to etch a wide variety of polymers as well as for [[sample cleaning]] and [[surface activation]].
 +
 
 +
==Ion milling==
 +
{{main|Ion milling}}
 +
 
 +
Ion milling uses an ion beam (typically [[argon]]) to [[sputtering|sputter]] material from the surface of the sample. It has very low selectivity (typically 1:1) but can be used on any material, even [[inert metal]]s.
 +
 
 +
===Nanoquest II Ion Mill===
 +
{{main|Nanoquest II Ion Mill}}
 +
The Nanoquest II Ion Mill is a 22 cm ion source for milling up to 150 mm wafers. It can etch most materials including metals, ceramics, semiconductors, dielectrics, etc.
  
 
==Method of operation==
 
==Method of operation==
Coming back to this...
+
{{see also|Reactive ion etching#Method of operation}}
  
 
==Parameters==
 
==Parameters==
<excerpted from RIE so I can reference it while writing the page>
+
Parameters that affect plasma etching characteristics include pressure, gas composition, and, [[plasma generation]] method, and generator power.
Many of the same parameters used in [[Plasma etching#Parameters|plasma etching]] apply to RIE, including pressure, gas composition, and generator power.  Of particular importance is the [[plasma generation]] method (commonly a parallel plate or ICP configuration), as they have different advantages depending on the material being etched.
 
  
 
===Gas composition===
 
===Gas composition===
Line 32: Line 62:
  
 
===Sample temperature===
 
===Sample temperature===
 +
Some plasma etching systems allow for the temperature of the sample to be controlled.  For most processes, the sample is kept between 20-50°C and are actively cooled to prevent damage to the mask, which is often [[photoresist]].  A liquid nitrogen cooled chuck, such as on the [[Oxford ICP RIE]] can be used to perform [[cryogenic etching]].  Also, some etch processes (e.g. [[polymer]] etching) benefit from a heated chuck (can be up to 200°C).
  
==Reactive ion etching==
+
==Complete tool list==
{{main|Reactive ion etching}}
+
For a complete list of plasma etching equipment available at the LNF, please see [[:Category:Plasma etching equipment|list of plasma etching equipment]] or the specific plasma etching category, above.
 
 
Reactive ion etching (RIE) is one of the most common forms of [[plasma etching]].  It typically uses a combination of chemically reactive elements and energetic ions to etch the desired material.  One major advantage to RIE over other forms of etching is that the process can be designed to be highly [[Isotropy|anisotropic]], allowing for much finer resolution and higher aspect ratios.
 
 
 
==Plasma ashing==
 
{{main|Plasma ashing}}
 
  
Plasma ashing typically refers to the removal of organics, particularly [[photoresist]] from a sample using a plasma discharge.  These processes typically use [[oxygen]] as the main etch gas and sometimes require a high temperature to enhance the reactivity.
+
<categorytree mode=pages>Plasma etching equipment</categorytree>
 
 
==Ion milling==
 
{{main|Ion milling}}
 
 
 
<stuff about ion milling> The LNF currently does not have any ion milling capabilities
 
 
 
==Equipment==
 
For a complete list of plasma etching equipment available at the LNF, please see [[:Category:Plasma etching equipment|list of plasma etching equipment]] or the specific plasma etching category, above.
 
  
 
==See also==
 
==See also==
 
* [[Plasma processing]]
 
* [[Plasma processing]]
  
 +
<!--
 
==Notes==
 
==Notes==
 
<footnotes>
 
<footnotes>
Line 59: Line 78:
 
==References==
 
==References==
 
<citations>
 
<citations>
 +
-->
  
 
==Further reading==
 
==Further reading==
* [https://docs.google.com/a/lnf.umich.edu/file/d/0B76AgohVTgqdamFiaE1mRk9mVmM/preview RIE Workshop January 16th, 2015]
+
* [http://lnf-wiki.eecs.umich.edu/wiki/User_Resources#LNF_Tech_Talks_.28technology_seminar_series.29 LNF Tech Talk for RIE]
  
 
[[Category:Plasma etching| ]]
 
[[Category:Plasma etching| ]]
 
[[Category:Plasma processing|Etching]]
 
[[Category:Plasma processing|Etching]]

Latest revision as of 09:40, 24 September 2021

Plasma etching
Rieoperation.svg
Technology Details
Other Names Dry etching
Technology Etching
Equipment List of plasma etching equipment

Plasma etching is a form of plasma processing designed to remove material from a sample using plasma discharges. It is highly controllable and can be used to etch a wide variety of materials. The most commonly used form of plasma etching is referred to in the microfabrication world as reactive ion etching (RIE). However, there are other types of plasma etching, including plasma ashing and ion milling. For a detailed overview of plasma etching in the LNF, please review the Tech Talk

Reactive ion etching

Main article: Reactive ion etching

Reactive ion etching (RIE) is one of the most common forms of plasma etching. It typically uses a combination of chemically reactive elements and energetic ions to etch the desired material. One major advantage to RIE over other forms of etching is that the process can be designed to be highly anisotropic, allowing for much finer resolution and higher aspect ratios.


P5000 RIE

Main article: P5000 RIE

The P5000 is a 3 chamber tool designed for production etching. Chambers A and B are configured for SiO2 and Si3N4 etching, where as chamber C is configured for polysilicon and amorphous silicon etching. Chambers B and C are restricted to CMOS clean devices where as chamber A is open to semi-clean devices.

STS APS DGRIE

Main article: STS Glass Etcher

The glass etcher excels at deep etching of fused silica but it also has a nearly vertical SiO2 etch. It's capable of very high bias powers enabling it to process difficult to etch materials.

LAM 9400 SE

Main article: LAM 9400

The LAM 9400 SE is an ICP etcher configured with a wide range of gas chemistries. It is mainly used to etch polysilicon but can also etch SiO2, Si3N4, compound semiconductors, some metals, and organic materials.

Oxford Plasmalab System 100

The oxford is another ICP etcher that also has a cryogenic chuck. By cooling the sample down to -150°C nearly vertical etches are possible in certain materials.

Plasmatherm 790

Main article: Plasmatherm 790

The Plasmatherm is configured with a variety of gases so that it can etch a wide array of materials. Most recipes tend to have slow etch rates on the order of 200 Å/min which is ideal for very thin films. The tool also has few material restrictions to allow it to process as many things as possible.

Deep reactive ion etching

Deep reactive ion etching (DRIE) is a term used to describe RIE processes that are designed to achieve high (>10:1) aspect ratios or etch depths greater than several microns. One common process is the Bosch process.

STS Pegasus

Main articles: STS Pegasus 4 and STS Pegasus 6

The STS Pegasus is a tool that utilizes the Bosch process for high aspect ratio etching of silicon. It uses SF6 for the etch step and C4F8 for passivation. It can achieve etch rates of up to 20 μm/min and aspect ratios up to 50:1.

STS Glass Etcher

Main article: STS Glass Etcher

The STS Glass Etcher is a DGRIE tool for high aspect ratio etching of silicon dioxide, glass, and fused silica.

Oxford ICP

The Oxford Plasmalab is an ICP RIE system capable of cryogenic etching. While not configured to etch silicon, it can etch compound semiconductors and perform a cryogenic etch of polymers.

Plasma ashing

Main article: Plasma ashing

Plasma ashing typically refers to the removal of organics, particularly photoresist from a sample using a plasma discharge. These processes typically use oxygen as the main etch gas and sometimes require a high temperature to enhance the reactivity.

YES-CV200RFS(E)

The YES-CV200RFS(E) (YES Plasma Stripper) is the primary plasma ashing equipment in the main cleanroom of the LNF. Its primary uses are low-temperature plasma descum and high-temperature stripping of photoresist but can also be used to etch a wide variety of polymers as well as for sample cleaning and surface activation.

Ion milling

Main article: Ion milling

Ion milling uses an ion beam (typically argon) to sputter material from the surface of the sample. It has very low selectivity (typically 1:1) but can be used on any material, even inert metals.

Nanoquest II Ion Mill

Main article: Nanoquest II Ion Mill

The Nanoquest II Ion Mill is a 22 cm ion source for milling up to 150 mm wafers. It can etch most materials including metals, ceramics, semiconductors, dielectrics, etc.

Method of operation

Parameters

Parameters that affect plasma etching characteristics include pressure, gas composition, and, plasma generation method, and generator power.

Gas composition

Main article: Plasma etching/Gases

The type and ratio of gases used in a plasma etch is chosen depending on the material being etched, the masking material, and the etch stop material in order to achieve high selectivity.

Pressure

Chamber pressure varies depending on the system and material being etched but typically ranges from 5 mTorr to 300 mTorr. Typically, more physical etches and etches designed to be very vertical or to have high aspect ratios run at lower pressures, while more reactive etches will use a higher pressure to increase the density of the reactive gases. Most plasma etching systems control the pressure in the chamber using a throttle valve on the exhaust port, allowing it to be accurately set regardless of the gases chosen.

Plasma source and power

The plasma generation source is critical to the function of the etch. Capacitively and inductively coupled RF plasmas are very common, particularly in RIE, but certain applications may use microwave sources, ECR sources, etc. Additionally, while in RIE the sample is typically placed directly under the source, sometimes the sample may be placed more indirectly from the source, such as in plasma ashing.

Sample temperature

Some plasma etching systems allow for the temperature of the sample to be controlled. For most processes, the sample is kept between 20-50°C and are actively cooled to prevent damage to the mask, which is often photoresist. A liquid nitrogen cooled chuck, such as on the Oxford ICP RIE can be used to perform cryogenic etching. Also, some etch processes (e.g. polymer etching) benefit from a heated chuck (can be up to 200°C).

Complete tool list

For a complete list of plasma etching equipment available at the LNF, please see list of plasma etching equipment or the specific plasma etching category, above.

See also


Further reading