Difference between revisions of "Plasmatherm 790/Processes/m pary1"

From LNF Wiki
Jump to navigation Jump to search
 
(2 intermediate revisions by the same user not shown)
Line 1: Line 1:
<!-- Make sure to add any other relevant categories -->
+
m_pary1 is an O<sub>2</sub> etch that can be used for parylene and other polymers.
 
[[Category:Processes]]
 
[[Category:Processes]]
{{warning|This page has not been released yet.}}
 
 
{{Infobox process
 
{{Infobox process
 
|image      =  
 
|image      =  
Line 14: Line 13:
 
|authors    =  
 
|authors    =  
 
}}
 
}}
m_pary1 is an O<sub>2</sub> etch that can be used for parylene and other polymers.
 
  
 
==Parameters==
 
==Parameters==
Line 37: Line 35:
 
*[[Parylene]] Etch Rate: 2000 Å/min
 
*[[Parylene]] Etch Rate: 2000 Å/min
 
* [[SPR 220]]: 2000 Å/min
 
* [[SPR 220]]: 2000 Å/min
* [[PDMS]]: XXX Å/min
 
  
 
===Mask Selectivity===
 
===Mask Selectivity===
 
* [[Silicon|Si]] Etch Rate: <5 Å/min
 
* [[Silicon|Si]] Etch Rate: <5 Å/min
 
* [[Silicon Dioxide|SiO2]] Etch Rate: <5 Å/min
 
* [[Silicon Dioxide|SiO2]] Etch Rate: <5 Å/min
 
===Sidewall Profile===
 
* Undercut: XXX
 
  
 
==Limitations==
 
==Limitations==
????
+
The limitations for this recipe are still being defined.
 
 
==Qualification==
 
<!-- Insert a description of how the process is qualified, generally this should include a chart to show the processes stability over time-->
 
Once a month m_pary1 is run on a XXX.  This etch rate is reported in the chart below.
 
 
 
{{#widget:Iframe
 
|url=https://docs.google.com/spreadsheets/d/1jDSV00jZ-1vppt2MgGRAJyXRvxEE3rJIlXtngODM9-0/pubchart?oid=1193338187&format=interactive
 
|width=700
 
|height=350
 
|border=0
 
}}
 
 
 
Then occasionally the recipe is tested on other materials as shown below.
 
{{#widget:Iframe
 
|url=https://docs.google.com/spreadsheets/d/1jDSV00jZ-1vppt2MgGRAJyXRvxEE3rJIlXtngODM9-0/pubchart?oid=1556592131&format=interactive
 
|width=700
 
|height=350
 
|border=0
 
}}
 

Latest revision as of 11:03, 18 August 2021

m_pary1 is an O2 etch that can be used for parylene and other polymers.

About this Process
Process Details
Equipment Plasmatherm 790
Technology RIE
Material Parylene
Mask Materials Photoresist
Gases Used O2



Parameters

Parameter Etch
RF Power 105 W
Pressure 100 mTorr
O2 Flow 99 sccm

Capabilities

Etch Rate

Mask Selectivity

  • Si Etch Rate: <5 Å/min
  • SiO2 Etch Rate: <5 Å/min

Limitations

The limitations for this recipe are still being defined.