Difference between revisions of "SSEC Wafer and Mask Cleaner"

From LNF Wiki
Jump to navigation Jump to search
 
(3 intermediate revisions by 2 users not shown)
Line 9: Line 9:
 
|materials =  
 
|materials =  
 
|mask = 5" and 7"  
 
|mask = 5" and 7"  
|size = wafer size 4" and 6"
+
|size = wafer size 6", 4" and mounted pieces.
 
|chemicals =NH<sub>4</sub>OH, HF, H<sub>2</sub>SO<sub>4</sub>, H<sub>2</sub>O<sub>2</sub>
 
|chemicals =NH<sub>4</sub>OH, HF, H<sub>2</sub>SO<sub>4</sub>, H<sub>2</sub>O<sub>2</sub>
 
|gases = na
 
|gases = na
Line 15: Line 15:
 
|sop = [https://docs.google.com/document/d/1DFrqq_0csijzsBPrTbcaNF7pCTEgiBVYpH14S369Ay0/preview SOP]
 
|sop = [https://docs.google.com/document/d/1DFrqq_0csijzsBPrTbcaNF7pCTEgiBVYpH14S369Ay0/preview SOP]
 
|processes = [[{{PAGENAME}}/Processes|Supported Processes]]
 
|processes = [[{{PAGENAME}}/Processes|Supported Processes]]
|userprocesses = [[LNF_User:{{PAGENAME}}_user_processes|User Processes]]
+
<!--|userprocesses = [[LNF_User:{{PAGENAME}}_user_processes|User Processes]] -->
 
|maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]]
 
|maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]]
 
}}
 
}}
  
{{warning|This page has not been released yet.}}
 
  
 
<!-- Insert the tool description here -->
 
<!-- Insert the tool description here -->
Line 25: Line 24:
  
 
==Announcements==
 
==Announcements==
*Update this with announcements as necessary
 
  
 
==Capabilities==
 
==Capabilities==

Latest revision as of 16:27, 18 October 2021

SSEC Wafer and Mask Cleaner
180011.jpg
SSEC Trilennium Wafer and Mask Cleaner
Equipment Details
Technology Wafer cleaning
Materials Restriction Metals
Mask Materials 5" and 7"
Sample Size wafer size 6", 4" and mounted pieces.
Chemicals Used NH4OH, HF, H2SO4, H2O2
Gases Used na
Equipment Manual
Overview System Overview
Operating Procedure SOP
Supported Processes Supported Processes
Maintenance Maintenance



The SSEC Wafer and Mask Cleaner is a...

Announcements

Capabilities

The SSEC Trilennium is a Photomask and single wafer cleaning system. This tool is used for post mechanical process cleaning of 100mm 150mm wafers, bonded small pieces, as well as 7", and 5" photolithography masks. The SSEC has four cleaning modules which include a single-brush scrub, 1.5 MHz quartz-enclosed megasonic transducer, a high-pressure scrub/rinse nozzle, and a multi-function dispense and etch nozzle. It is also equipped with a selection of mixing and spray dispensers. The SSES's four dispense arms offer access to different internal chemistry for sample cleans. Software supported recipes with chemical dilution, sample rinse and spin dry cycles allow for safe and repeatable operation reducing hazard and exposure to processors.

System overview

Hardware details

  • 4", 6", wafer chuck and 5" mask chucks
  • Chuck adapter for Laper mounting glass
  • Rotary PVA brush arm
  • Spray arm
  • Megasonic arm
  • HF arm
  • top and bottom DI rinsing
  • high rpm for sample drying
  • Exchangeable chucks

Substrate requirements

  • Wafer Size 4" and 6"
  • Sample can be mounted on a wafer
  • mounting wax cannot be exposed to the chemicals
  • variable wafer height

Material restrictions

Wafers with exposed mounting materials that can react with the chemicals in the tool are not allowed. There is direct contact between the PVA brush and the surface, cleaning is done by scrubbing, therefore delicate structures should not be exposed to the brush or the megasonic arm. All materials that are not damaged by or react with the chemicals in the tool are accepted

Supported processes

There are several processes for this tool supported by the LNF, which are described in more detail on the Processes page.

  • Wafer (4" and 6" ) cleaning after Si and oxide CMP
  • Piranha Mask cleaning
  • Pieces Processing is available when mounted on a wafer

Standard operating procedure

Widget text will go here.

Checkout procedure

Training and check out of the SSEC is done typically completed together when processing on the CMP

  1. Read through the Standard Operating Procedure.
  2. Create a Helpdesk Ticket requesting training.
  3. A tool engineer will schedule a time for initial training.
  4. Practice with your mentor or another authorized user until you are comfortable with tool operation.
  5. Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool.

Maintenance

Process name

File:Pilars pic.jpg