Difference between revisions of "STS Pegasus 4/Processes/LNF Recipe 2"
Line 1: | Line 1: | ||
− | [[Category:Processes]] | + | [[Category:Processes]]{{Infobox process |
− | {{Infobox process | ||
|image = STS Pegasus_4_Recipe_2_2_um.jpg | |image = STS Pegasus_4_Recipe_2_2_um.jpg | ||
|caption = Recipe 2 for 20 min, 2 µm Trench | |caption = Recipe 2 for 20 min, 2 µm Trench | ||
Line 12: | Line 11: | ||
}} | }} | ||
− | + | '''{{SUBPAGENAME}}''' is designed for deep and through wafer etching of small (1-500 μm) features on masks with a low exposed area (<20%). Undercut and scalloping is larger than LNF Recipe 1, but etch rates are higher (4-13 μm/min). | |
==Procedure== | ==Procedure== | ||
Line 115: | Line 114: | ||
===Mask Selectivity=== | ===Mask Selectivity=== | ||
* Photoresist Etch Rate | * Photoresist Etch Rate | ||
− | ** [[SPR 220]]: | + | ** [[SPR 220]]: 400 Å/min |
* [[SiO2|SiO2]] Etch Rate | * [[SiO2|SiO2]] Etch Rate | ||
** Thermal oxide: 155 Å/min | ** Thermal oxide: 155 Å/min | ||
Line 121: | Line 120: | ||
** LTO: 185 Å/min | ** LTO: 185 Å/min | ||
** PECVD: unknown | ** PECVD: unknown | ||
+ | * [[Silicon nitride]] Etch Rate | ||
+ | ** LPCVD Nitride: 168 Å/min | ||
+ | ** LPCVD Low Stress Nitride: 108 Å/min | ||
+ | |||
===Sidewall Profile=== | ===Sidewall Profile=== |
Revision as of 16:23, 1 March 2021
About this Process | |
---|---|
Recipe 2 for 20 min, 2 µm Trench | |
Process Details | |
Equipment | STS Pegasus 4 |
Technology | DRIE |
Material | Si |
Mask Materials | PR, SiO2 |
Gases Used | SF6, C4F8, O2 |
Date Created | Jan 1, 2009 |
Authored By | SPTS, Brian Vanderelzen, Kevin Owen |
LNF Recipe 2 is designed for deep and through wafer etching of small (1-500 μm) features on masks with a low exposed area (<20%). Undercut and scalloping is larger than LNF Recipe 1, but etch rates are higher (4-13 μm/min).
Contents
Procedure
Run LNF Recipe 2 for up to 20 minutes with the 1 minute automated clean recipe, longer etches should be run with the 3 minute automated clean recipe.
Parameters
Parameter | Dep | Etch | |
---|---|---|---|
Boost | Main | ||
ICP Power | 2000 W | 2800 W | |
Bias Power | 0 W | 100 W | 40 W |
Bias Pulsing | N/A | 40 Hz 80% | |
Pressure | 25 mTorr | 30 mTorr | 100 mTorr |
C4F8 Flow | 200 sccm | 0 sccm | |
SF6 Flow | 0 sccm | 450 sccm | |
O2 Flow | 0 sccm | 45 sccm | |
Time | 4 s | 1.5 s | 7 s |
Chuck Temperature | 20°C |
Capabilities
This process, as is true with all Bosch processes, varies depending on feature size and density, due to RIE lag. The below table contains typical results for 2 µm, 10 µm, and 100 µm trenches and the following sections expand on these results. It is strongly recommended that the process is tested with any new mask.
Parameter | 2 µm Trench | 10 µm Trench | 100 µm Trench |
---|---|---|---|
Etch Rate | 6.8 µm/min | 10.4 µm/min | 14.4 µm/min |
Undercut | 0.85 µm | 0.9 µm | |
Scallop height | 1.6 µm | 2.6 µm | 3 µm |
Scallop depth | 0.65 µm | 0.74 µm | |
Sidewall angle | 89.8° | 89.8° | 88.4° |
Etch Rate
The etch rate will vary with the size of the trench being etched. Larger trenches tend to etch faster, this trend can be seen below.
Mask Selectivity
- Photoresist Etch Rate
- SPR 220: 400 Å/min
- SiO2 Etch Rate
- Thermal oxide: 155 Å/min
- HTO: 175 Å/min
- LTO: 185 Å/min
- PECVD: unknown
- Silicon nitride Etch Rate
- LPCVD Nitride: 168 Å/min
- LPCVD Low Stress Nitride: 108 Å/min
Sidewall Profile
Mask undercut, scalloping, and sidewall angle are shown below.
Trench
Running recipe 2 for 20 minutes with 3 µm of SPR 220 on a silicon wafer with our standard DRIE characterization mask can produce a profile like this.
Limitations
LNF Recipe 2 is known to work under a range of conditions. Outside of these conditions, etch performance is not guaranteed. Etch rates will vary depending on feature size, etch length, open area, etc. We strongly recommend testing the process prior to running device wafers.
One common problem seen when running outside of these conditions is grassing. This occurs when the C4F8/SF6 ratio is unbalanced causing micromasking. This creates tiny pillars of un-etched Si and teflon on the surface of the feature. An example of grassing can be seen below.
Etch Length
- Generally 30 sec - 20 minutes is known to work well.
- Recipe 2 is tested for 10 min monthly to verify performance. Longer etches may cause undesirable results.
Feature Size
- Tested on features from 1 μm - 100 μm.
- The standard characterization pattern used to qualify Recipe 2 has open lines, trenches, and grating patterns from 1 μm - 100 μm. Larger and smaller features have not been tested and may not provide the expected results.
Aspect ratio
This section's factual accuracy is disputed. |
- Aspect ratio limitations vary depending on feature size. For instance, 2 μm features have an aspect ratio limit of around 50:1. For larger features, aspect ratio is typically limited to etch length (see above).
Open area
- Very large features and masks with total open area >20% are known to degrade etch performance and should be tested prior to running device wafers. For masks with more than 20% open area Recipe 3 may provide better results.
Multi-step etching
- Multi-step processing (i.e. running two or more etches with overlapping features) with the Bosch process is known to cause undesirable results. Please contact the tool engineers if you wish to run such a process so we can help reduce or eliminate issues.
Qualification
Every third Friday Recipe 2 is run for 10 minutes on a Si wafer with our standard DRIE characterization mask, with 3 µm of SPR 220. This is performed after the weekly clean, and a 100 µm trench is measured in 5 locations with the Zygo NewView 5000. This etch rate is reported in the chart below.