Difference between revisions of "STS Pegasus 6"

From LNF Wiki
Jump to navigation Jump to search
Line 1: Line 1:
{{#vardefine:toolid|14031}}{{#vardefine:technology|DRIE}}{{#vardefine:restriction|2}}
+
{{#vardefine:toolid|14031}}{{#vardefine:technology|DRIE}}{{#vardefine:restriction|2}}{{infobox equipment
{{infobox equipment
 
 
|manufacturer = [http://www.spts.com/ SPTS Technologies Ltd.]
 
|manufacturer = [http://www.spts.com/ SPTS Technologies Ltd.]
 
|model = Pegasus
 
|model = Pegasus
Line 7: Line 6:
 
|size = 150 mm
 
|size = 150 mm
 
|gases = [[Ar|Ar]], [[C4F8|C<sub>4</sub>F<sub>8</sub>]], [[O2|O<sub>2</sub>]], [[SF6|SF<sub>6</sub>]]
 
|gases = [[Ar|Ar]], [[C4F8|C<sub>4</sub>F<sub>8</sub>]], [[O2|O<sub>2</sub>]], [[SF6|SF<sub>6</sub>]]
|overview = [https://docs.google.com/document/d/1y4T8TsG121BpsmMuF4xN5vlqkvXK5AOLQUzZWGPn51U/preview System Overview]
+
|overview = [[{{PAGENAME}}#System overview | System overview]]
|sop = [https://docs.google.com/a/lnf.umich.edu/document/d/10E3Ez04ZIe0-68K5jSJCjzB26l2UW0vUySu4fgbQMXQ/preview SOP]
+
|sop = [https://docs.google.com/a/lnf.umich.edu/document/d/1cvr1FTl9msRafNlV92n2P8mCiYxMw1PMev2LYvjNtzs/preview User manual]
 
|processes = [[{{PAGENAME}}/Processes|Supported Processes]]
 
|processes = [[{{PAGENAME}}/Processes|Supported Processes]]
|userprocesses = [[LNF_User:{{PAGENAME}}_User_Processes|User Processes]]
+
|userprocesses = [[LNF_User:{{PAGENAME}}_User_Processes|User processes]]
 
|maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]]
 
|maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]]
 
}}
 
}}
The SPTS Pegasus 6 is a [[DRIE|Deep reactive ion etch (DRIE)]] tool used to etch 6" (150 mm) silicon wafers using the Bosch process.  It has a high power [[Plasma processing/Inductively coupled plasma|ICP]] source and is capable of fast (over 20 μm/min), high aspect ratio (up to 100:1) [[Isotropy|anisotropic]] etching of silicon.  Fast-acting, high precision [[MFC]]s and [[throttle valve|VAT pendulum valve]] and a high capacity [[Vacuum pump#Turbomolecular pump|turbo pump]] provide fast switching times (down to 1s), enabling reduced undercut and sidewall roughness.  Sub-micron feature etching has been demonstrated (down to 100 nm linewidth).  Etch rate varies with feature size and density, so a characterization run is strongly recommended with any new mask.
+
The [[{{PAGENAME}}]] is a [[DRIE|Deep Reactive Ion Etch (DRIE)]] tool manufactured by [http://www.spts.com/ SPTS Technologies]. It is used for fast, high aspect ratio anisotropic etching of 6" (150 mm) [[silicon]] wafers using the [https://patents.google.com/patent/US5501893A/en Bosch] process.  It is also capable of slow, controlled etching with minimal undercut, sidewall roughness, and notching; sub-micron feature etching has been demonstrated down to 100 nm linewidth.  Etch rate varies with feature size and density, so a characterization run is strongly recommended with any new mask.
  
 
==Announcements==
 
==Announcements==
Line 26: Line 25:
 
* High selectivity to photoresist and [[SiO2|SiO<sub>2</sub>]]
 
* High selectivity to photoresist and [[SiO2|SiO<sub>2</sub>]]
  
==System Overview==
+
==System overview==
===Hardware Details===
+
===Hardware details===
 
*Gases
 
*Gases
 
** High Precision Fast Acting Chamber Mounted MFC's
 
** High Precision Fast Acting Chamber Mounted MFC's
Line 55: Line 54:
  
 
===Substrate Requirements===
 
===Substrate Requirements===
* 150 mm (6") wafers
+
The {{PAGENAME}} is equipped to handle 6” (150 mm wafers) up to 3 mm thick. All smaller samples must be mounted to a carrier wafer. All standard major and minor flats and full-round wafers are allowed.
** Pieces and smaller wafers can be mounted to a carrier wafer at the [[Sample Mounting Station]]
+
 
* If the etch will leave less than 100 µm remaining, a carrier wafer ''must'' be used
+
====Sample mounting====
* Sample/carrier must be conductive, or a semiconductor for wafer to clamp on electrostatic chuck
+
{{see also | Sample mounting}}
** [[Pyrex]] glass with an Si layer on top has worked
+
 
 +
If your sample and process meet any of the following criteria, you MUST mount your sample to a 4” carrier wafer:
 +
 
 +
* Sample size is smaller than a 6” (150 mm) wafer
 +
* Substrate material is non-conducting (e.g. glass)
 +
* Sample thickness is less than 100 μm
 +
* Remaining thickness after etch is less than 100 μm
 +
{{note|Etch depth must be calculated for the deepest feature. An etch rate test should be performed to determine this for each process and mask design.|reminder}}
 +
 
 +
For all standard (Bosch process) recipes, an SiO2 coated carrier is required, unless otherwise instructed by a tool engineer.  For non-bosch process recipes, create a helpdesk ticket and staff will help you determine the appropriate material.
 +
 
  
 
===Material Restrictions===
 
===Material Restrictions===
{{material restrictions}}
+
[[File:Etch_material_layers.png|right|diagram of typical material stack during an etch]]
 +
The {{PAGENAME}} is designated as a [[{{#switch: {{#var:restriction}}
 +
| 1 = CMOS Clean
 +
| 2 = Semi-Clean
 +
| 3 = Metals
 +
| 4 = General
 +
| Undefined
 +
}}]] class tool. A full list of approved materials is included at the end of this section. In addition to the restrictions in this list, materials can be classified into four categories, detailed below: materials that may be '''etched''', materials that can be used as '''masks''', '''etch stop''' materials, and '''buried''' materials. Use of any material outside of these conditions requires approval by the LNF staff via a helpdesk ticket.
 +
 
 +
====Materials etched====
 +
The STS Pegasus 4 is intended for etching single crystal [[silicon]]. While it is capable of etching silicon dioxide, silicon nitride, and poly-silicon, there are other tools at the LNF with better performance for these materials. There are certain restrictions for where these materials may be deposited, as detailed in the approved materials list.
 +
 
 +
====Mask materials====
 +
This includes any material that will be exposed to the plasma for the majority of the process. The most common mask materials are [[photoresist]] and [[silicon dioxide]]. [[Silicon nitride]] is also allowed, although it has not been characterized. There are certain restrictions for where these materials may be deposited, as detailed in the approved materials list.
 +
 
 +
====Etch stop materials====
 +
This includes any material that will be exposed briefly to the plasma. All materials listed in the approved materials list on the wiki are allowed, including approved mounting materials.
 +
 
 +
====Buried materials====
 +
These materials may be present on the sample, but may not be exposed to the plasma. They may be covered by the mask or on the back of the sample, provided that the sample is mounted to a carrier wafer. Materials listed in the approved materials list on the wiki are allowed.
 +
 
 +
====Approved materials====
 +
Below is a list of approved materials for the tool. ''Approved'' means the material is allowed in the tool under the conditions described above. If a material is not listed, please create a helpdesk ticket or email [mailto:info@umich.edu info@lnf.umich.edu] for any material requests or questions.
 +
 
 +
<div class="NavFrame" style="font-size:100%;margin:0;border-style:none;padding:0;">
 +
  <div class="NavFrame" style="border-style:solid;background:#fff;padding:0.2em;">
 +
    <div class="NavHead" style="background:#ddd;text-align:center;">{{Big|Approved materials}}</div>
 +
    <div class="NavContent" style="text-align:left;">{{#widget:ApprovedMaterials|toolid={{#var:toolid}}{{{1|}}}|header=N}}</div>
 +
  </div>
 +
</div>
  
 
==Supported Processes==  
 
==Supported Processes==  
Line 72: Line 110:
  
 
==Standard Operating Procedure==
 
==Standard Operating Procedure==
{{#widget:GoogleDoc|key=10E3Ez04ZIe0-68K5jSJCjzB26l2UW0vUySu4fgbQMXQ}}
+
'''[https://drive.google.com/file/d/19IFXiNNWGF2U1n90bp0SGgTK1GLJXyFt/view?usp=sharing PDF Copy]'''
 +
 
 +
{{#widget:GoogleDoc|key=1cvr1FTl9msRafNlV92n2P8mCiYxMw1PMev2LYvjNtzs}}
  
 
==Checkout Procedure==
 
==Checkout Procedure==
<!-- Describe the checkout procedure for the tool. For example: -->
+
Follow this procedure to receive authorization to run supported processes. Practicing with an authorized user or staff member prior to checkout is strongly encouraged.
# Complete the [[Sample_mounting#Training_modules|sample mounting course]]. If you have already completed this for another tool, you do not need to complete it again.
+
 
# Read through the Standard Operating Procedure above.
+
# Complete the [[Sample_mounting#Training_modules|sample mounting course]]. If you have already completed this for another tool, you do not need to complete it again.
 +
# Read through the User Manual above.
 +
# Accurately complete the [https://docs.google.com/forms/d/e/1FAIpQLScckma-M-SW0B3ZiY_OpIHgq3lk445RuX1WryRefSkzmzh1Jg/viewform checkout quiz]. You may retake as necessary until all answers are correct.
 
# Complete the [https://docs.google.com/forms/d/1dbMYfNd4zqoT2h7xhpfFVNnMwb8EXUaWjqs8JZtpRjY/viewform training request form].
 
# Complete the [https://docs.google.com/forms/d/1dbMYfNd4zqoT2h7xhpfFVNnMwb8EXUaWjqs8JZtpRjY/viewform training request form].
# Create a [http://ssel-sched.eecs.umich.edu/sselScheduler/ResourceContact.aspx?tabindex=3&path=0:0:0:{{#var:toolid}} Helpdesk Ticket] requesting training.
+
# Create a [http://ssel-sched.eecs.umich.edu/sselScheduler/ResourceContact.aspx?tabindex=3&path=0:0:0:{{#var:toolid}} helpdesk ticket] requesting a checkout session.
# A tool engineer will schedule a time for initial training.
+
# Authorization will be provided pending successful completion of the quiz and demonstration of proper tool use in the presence of a tool engineer.
# Practice with your mentor or another authorized user until you are comfortable with tool operation.
 
<!--# Complete the SOP quiz [<link> here].-->
 
# Schedule a checkout session with a tool engineer via the helpdesk ticket system. If this checkout is successful, the engineer will authorize you on the tool.
 
  
 
==Maintenance==  
 
==Maintenance==  

Revision as of 15:23, 8 April 2021

STS Pegasus 6
14031.jpg
Equipment Details
Manufacturer SPTS Technologies Ltd.
Model Pegasus
Technology DRIE
Materials Restriction Semi-Clean
Material Processed Si
Mask Materials PR, SiO2
Sample Size 150 mm
Gases Used Ar, C4F8, O2, SF6
Equipment Manual
Overview System overview
Operating Procedure User manual
Supported Processes Supported Processes
User Processes User processes
Maintenance Maintenance


The STS Pegasus 6 is a Deep Reactive Ion Etch (DRIE) tool manufactured by SPTS Technologies. It is used for fast, high aspect ratio anisotropic etching of 6" (150 mm) silicon wafers using the Bosch process. It is also capable of slow, controlled etching with minimal undercut, sidewall roughness, and notching; sub-micron feature etching has been demonstrated down to 100 nm linewidth. Etch rate varies with feature size and density, so a characterization run is strongly recommended with any new mask.

Announcements

    • [2017-11-10 - Annual maintenance
    • LNF Recipe 1: 6.67 um/min
    • LNF Recipe 2: 16.38 um/min
    • LNF Recipe 3: 17.70 um/min

Capabilities

  • High etch rates ~15 µm /min
  • Nearly vertical sidewalls ~89°
  • High selectivity to photoresist and SiO2

System overview

Hardware details

  • Gases
    • High Precision Fast Acting Chamber Mounted MFC's
      • C4F8 - 400 sccm
      • SF6 - 1200 sccm
      • O2 - 200 sccm
    • Standard Gasbox Mounted MFC
      • Ar - 283 sccm
  • Pressure
    • 1550 L/sec Mag 2000 CTS Turbo Pump
    • 2-250 mTorr
    • Fast acting VAT pendulum valve
      • Allows fast switching of BOSCH recipes (~1.5 seconds minimum)
  • Chuck
    • 150 mm Wafer
    • 0-20 Torr Backside He Cooling
    • -20°C to 20°C
  • Chamber
    • 120°C Walls
    • Aluminum Walls
  • RF
    • 5 kW, 13.56 Mhz Coil
    • 300 W, 13.56 Mhz Platen
    • 500 W, 380 kHz Platen with Pulsed Power Supply
    • 10 A Chamber Outer Electromagnet
    • 30 A Chamber Outer Electromagnet

Substrate Requirements

The STS Pegasus 6 is equipped to handle 6” (150 mm wafers) up to 3 mm thick. All smaller samples must be mounted to a carrier wafer. All standard major and minor flats and full-round wafers are allowed.

Sample mounting

See also: Sample mounting

If your sample and process meet any of the following criteria, you MUST mount your sample to a 4” carrier wafer:

  • Sample size is smaller than a 6” (150 mm) wafer
  • Substrate material is non-conducting (e.g. glass)
  • Sample thickness is less than 100 μm
  • Remaining thickness after etch is less than 100 μm
Etch depth must be calculated for the deepest feature. An etch rate test should be performed to determine this for each process and mask design.

For all standard (Bosch process) recipes, an SiO2 coated carrier is required, unless otherwise instructed by a tool engineer. For non-bosch process recipes, create a helpdesk ticket and staff will help you determine the appropriate material.


Material Restrictions

diagram of typical material stack during an etch

The STS Pegasus 6 is designated as a Semi-Clean class tool. A full list of approved materials is included at the end of this section. In addition to the restrictions in this list, materials can be classified into four categories, detailed below: materials that may be etched, materials that can be used as masks, etch stop materials, and buried materials. Use of any material outside of these conditions requires approval by the LNF staff via a helpdesk ticket.

Materials etched

The STS Pegasus 4 is intended for etching single crystal silicon. While it is capable of etching silicon dioxide, silicon nitride, and poly-silicon, there are other tools at the LNF with better performance for these materials. There are certain restrictions for where these materials may be deposited, as detailed in the approved materials list.

Mask materials

This includes any material that will be exposed to the plasma for the majority of the process. The most common mask materials are photoresist and silicon dioxide. Silicon nitride is also allowed, although it has not been characterized. There are certain restrictions for where these materials may be deposited, as detailed in the approved materials list.

Etch stop materials

This includes any material that will be exposed briefly to the plasma. All materials listed in the approved materials list on the wiki are allowed, including approved mounting materials.

Buried materials

These materials may be present on the sample, but may not be exposed to the plasma. They may be covered by the mask or on the back of the sample, provided that the sample is mounted to a carrier wafer. Materials listed in the approved materials list on the wiki are allowed.

Approved materials

Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under the conditions described above. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.

Supported Processes

The primary recipes for the STS Pegasus 6 are used for deep silicon etching. There are four main Bosch process recipes with varying properties. The tool also has an oxynitride etch that may be used to etch a mask layer prior to the Si etch, but this may be performed by other tools more effectively. There is also a thinning recipe designed to quickly etch a wafer without a mask. More details on supported processes can be found on the Processes page.

In addition to these, this tool has a number of user-created recipes for specific processes. Some of these recipes are documented on STS Pegasus 6 User Processes. For more information, please contact the tool engineers via the helpdesk ticket system.

Standard Operating Procedure

PDF Copy

Widget text will go here.

Checkout Procedure

Follow this procedure to receive authorization to run supported processes. Practicing with an authorized user or staff member prior to checkout is strongly encouraged.

  1. Complete the sample mounting course. If you have already completed this for another tool, you do not need to complete it again.
  2. Read through the User Manual above.
  3. Accurately complete the checkout quiz. You may retake as necessary until all answers are correct.
  4. Complete the training request form.
  5. Create a helpdesk ticket requesting a checkout session.
  6. Authorization will be provided pending successful completion of the quiz and demonstration of proper tool use in the presence of a tool engineer.

Maintenance

In order to provide reliable operating conditions, maintenance is performed regularly on the tool including inspecting and cleaning the chamber. The following regular maintenance is performed on the STS Pegasus 6:

  • Weekly
    • General tool check
    • Qualification wafer run
    • Check wafer centering
  • Monthly
    • Wipe down chamber walls
    • Clean process kit, clamping ring
    • Plasma clean and condition
  • Annually
    • Change and clean source ceramics
  • Biennially
    • Refurbish turbo pump
    • Clean APC valve, foreline valve, baratron valve

After any chamber maintenance, the etch rate of the standard recipes is checked, as described below.

LNF Recipe 1, 2, 3

To verify the condition of the tool an etch is performed on recipe 1, 2 and 3, rotating each week. The etch rate is measured on a 100 μm wide trench in 5 locations. The average of this is shown below.