Difference between revisions of "Tempress S1T2 - LTO 4""

From LNF Wiki
Jump to navigation Jump to search
 
(18 intermediate revisions by 3 users not shown)
Line 8: Line 8:
 
|gases = [[Nitrogen|N<sub>2</sub>]], [[Oxygen|O<sub>2</sub>]], [[Silane|SiH<sub>4</sub>]]
 
|gases = [[Nitrogen|N<sub>2</sub>]], [[Oxygen|O<sub>2</sub>]], [[Silane|SiH<sub>4</sub>]]
 
|overview = [[{{PAGENAME}}#System_Overview | System Overview]]
 
|overview = [[{{PAGENAME}}#System_Overview | System Overview]]
|sop = [https://docs.google.com/document/d/1OwR52cfcxqd580bLlHpUAx3OpyCa3hjq9Q63d8lPoHA/edit SOP]
+
|sop = [https://docs.google.com/document/d/1OwR52cfcxqd580bLlHpUAx3OpyCa3hjq9Q63d8lPoHA/preview SOP]
 
|processes = [[LTO]]
 
|processes = [[LTO]]
 
|maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]]
 
|maintenance = [[{{PAGENAME}}#Maintenance | Maintenance]]
Line 15: Line 15:
  
 
==Announcements==
 
==Announcements==
* The tool is down until further notice.
+
* There are no announcements at this time.
  
 
==Capabilities==
 
==Capabilities==
* [[LTO]] - Low Temperature Oxide
+
* [[Silicon dioxide|Low Temperature Oxide]] - LTO
* Maximum Thickness - 12 µm
+
* Maximum Thickness - 2µm
  
 
==System Overview==
 
==System Overview==
Line 32: Line 32:
 
* Pieces up to 100 mm (4")
 
* Pieces up to 100 mm (4")
 
* Can process up to 25, 100 mm (4") wafers
 
* Can process up to 25, 100 mm (4") wafers
* All samples must go through the [[RCA Bench 81|RCA Pre-Furnace clean]] procedure before being processed in the furnace
+
{{note|All samples must go through the [[RCA Bench 81|RCA Pre-Furnace clean]] procedure before being processed in the furnace|reminder}}
  
 
===Material Restrictions===
 
===Material Restrictions===
Line 38: Line 38:
  
 
==Supported Processes==
 
==Supported Processes==
* [[LTO]] (Low Temperature Oxide)
+
* [[Silicon dioxide|LTO]] - Low temperature oxide
  
 
===Process Details===
 
===Process Details===
Line 44: Line 44:
  
 
===Characterization Data===
 
===Characterization Data===
* [https://docs.google.com/spreadsheets/d/1UUWHyJu2hqcF4dG7UPVhik6zyYSVGKUyvBuujMDwrVo/edit#gid=1363230001 LTO]
+
* [https://docs.google.com/spreadsheets/d/1UUWHyJu2hqcF4dG7UPVhik6zyYSVGKUyvBuujMDwrVo/edit#gid=1363230001 LTO - Low temperature oxide]
  
 
==Standard Operating Procedure==
 
==Standard Operating Procedure==
* [https://docs.google.com/document/d/1OwR52cfcxqd580bLlHpUAx3OpyCa3hjq9Q63d8lPoHA/preview Standard operating procedure]
+
* [https://docs.google.com/document/d/1OwR52cfcxqd580bLlHpUAx3OpyCa3hjq9Q63d8lPoHA/preview Quick reference SOP]
 +
* [https://docs.google.com/document/d/12iBrxud7Ge4Pw6b4us1srWDu0t42nBhhx96OBM5pmeg/edit?usp=sharing Detailed SOP]
  
 
==Checkout Procedure==
 
==Checkout Procedure==
# Read the Standard Operating Procedure above.
 
 
# Create a [http://ssel-sched.eecs.umich.edu/sselScheduler/ResourceContact.aspx?tabindex=3&path=0:0:0:{{#var:toolid}} Help desk Ticket] requesting training.
 
# Create a [http://ssel-sched.eecs.umich.edu/sselScheduler/ResourceContact.aspx?tabindex=3&path=0:0:0:{{#var:toolid}} Help desk Ticket] requesting training.
# A process engineer will schedule a time for initial training. After completing training the user can attempt a checkout session. It may be necessary for some users to participate in more than one training session, prior to attempting check out.
+
# LNF staff will schedule a time for initial training. After completing training the user can attempt a checkout session. It may be necessary for some users to participate in more than one training session, prior to attempting check out.
# Schedule a checkout session with a tool engineer via the [http://ssel-sched.eecs.umich.edu/sselScheduler/ResourceContact.aspx?tabindex=3&path=0:0:0:{{#var:toolid}} Help desk Ticket] system. If this checkout is successful, the engineer will authorize you on the tool.
+
# Schedule a checkout session with a tool engineer via the existing Helpdesk ticket. If this checkout is successful, the engineer will authorize you on the tool.
  
 
==Maintenance==  
 
==Maintenance==  
* Maintenance is performed on the tool, after 30-40 µm of accumulated deposition.  
+
* Maintenance is performed on the tool, after 30-40µm of accumulated deposition.  
 
* Maintenance consists of replacing the following quartz parts within the tube.  
 
* Maintenance consists of replacing the following quartz parts within the tube.  
 
** Cage
 
** Cage
 
** Twin rods
 
** Twin rods
 
** Bridge for twin rods
 
** Bridge for twin rods
** Thermo couple sheath
+
** Thermocouple sheath
 
** SiH<sub>4</sub> injector
 
** SiH<sub>4</sub> injector
 
** O<sub>2</sub> injector
 
** O<sub>2</sub> injector
 
** Dummy wafers
 
** Dummy wafers
 +
 +
* Tube is replaced at 100µm of accumulated deposition.

Latest revision as of 13:48, 11 January 2022

Tempress S1T2 - LTO 4"
30061.jpg
Equipment Details
Technology LPCVD
Materials Restriction CMOS Clean
Sample Size Pieces up to 100 mm (4")
Gases Used N2, O2, SiH4
Equipment Manual
Overview System Overview
Operating Procedure SOP
Supported Processes LTO
Maintenance Maintenance


The Tempress diffusion system is a modular horizontal furnace designed to process silicon wafers as part of the manufacturing technology of semiconductor, optical, MEMS, and solar devices. LTO is formed by using a process called LPCVD, Low Pressure Chemical Vapor Deposition. Chemical vapor deposition forms thin films on the surface of a substrate by thermal decomposition and/or reaction of gaseous compounds. The desired material is deposited directly from the gas phase onto the surface of the wafer. The important factors are gas ratio, temperature, and pressure.

Announcements

  • There are no announcements at this time.

Capabilities

System Overview

Hardware Details

  • N2 - Maximum flow 1 slpm
  • N2 - Maximum flow 10 slpm
  • O2 - Maximum flow 500 sccm
  • SiH4 - Maximum flow 100 sccm

Substrate Requirements

  • Silicon only
  • Pieces up to 100 mm (4")
  • Can process up to 25, 100 mm (4") wafers
All samples must go through the RCA Pre-Furnace clean procedure before being processed in the furnace

Material Restrictions

The Tempress S1T2 - LTO 4" is designated as a CMOS Clean class tool. Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.


Supported Processes

  • LTO - Low temperature oxide

Process Details

Characterization Data

Standard Operating Procedure

Checkout Procedure

  1. Create a Help desk Ticket requesting training.
  2. LNF staff will schedule a time for initial training. After completing training the user can attempt a checkout session. It may be necessary for some users to participate in more than one training session, prior to attempting check out.
  3. Schedule a checkout session with a tool engineer via the existing Helpdesk ticket. If this checkout is successful, the engineer will authorize you on the tool.

Maintenance

  • Maintenance is performed on the tool, after 30-40µm of accumulated deposition.
  • Maintenance consists of replacing the following quartz parts within the tube.
    • Cage
    • Twin rods
    • Bridge for twin rods
    • Thermocouple sheath
    • SiH4 injector
    • O2 injector
    • Dummy wafers
  • Tube is replaced at 100µm of accumulated deposition.