Tempress S6T4 - LTO 4"
The Tempress diffusion system is a modular horizontal furnace designed to process silicon wafers as part of the manufacturing technology of semiconductor, optical, MEMS, and solar devices. LTO is formed by using a process called LPCVD, Low Pressure Chemical Vapor Deposition. Chemical vapor deposition forms thin films on the surface of a substrate by thermal decomposition and/or reaction of gaseous compounds. The desired material is deposited directly from the gas phase onto the surface of the wafer. The important factors are gas ratio, temperature, and pressure.
| Tempress S6T4 - LTO 4" | |
|---|---|
| Equipment Details | |
| Technology | LPCVD |
| Materials Restriction | Broad |
| Sample Size | 100mm |
| Gases Used | N2, O2, SiH4 |
| Equipment Manual | |
| Overview | System Overview |
| Operating Procedure | SOP |
| Supported Processes | LTO |
| Maintenance | Maintenance |
Contents
Announcements
- The tool is down for a tube change.
Capabilities
- Low Temperature Oxide - LTO
- Maximum Thickness - 12µm
System overview
Hardware details
Substrate requirements
- Silicon only
- 100 mm (4")
- Can process up to 25, 100 mm (4") wafers
- A solvent clean must be performed on any samples prior to the run
- 5 min clean Acetone
- 5 min clean IPA
- 5 min fresh clean Acetone
- 5 min fresh clean IPA
- 3 min DI rinse
Material restrictions
Below is a list of approved materials for the tool. Approved means the material is allowed in the tool under normal circumstances. If a material is not listed, please create a helpdesk ticket or email info@lnf.umich.edu for any material requests or questions.
Supported processes
Standard operating procedure
Checkout procedure
- Create a Help desk Ticket requesting training.
- A process engineer will schedule a time for initial training. After completing training the user can attempt a checkout session. It may be necessary for some users to participate in more than one training session, prior to attempting check out.
- Schedule a checkout session with a tool engineer via the Help desk Ticket system. If this checkout is successful, the engineer will authorize you on the tool.
Maintenance
- Maintenance is performed on the tool, after 30-40µm of accumulated deposition.
- Maintenance consists of replacing the following quartz parts within the tube.
- Cage
- Twin rods
- Bridge for twin rods
- Thermocouple sheath
- SiH4 injector
- O2 injector
- Dummy wafers
- Tube is replaced at 100µm of accumulated deposition.